INSTRUCTIONS CKX41/CKX31 CULTURE MICROSCOPES This instruction manual is for the Olympus Culture Microscopes Models CKX41 and CKX31. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this microscope, we recommend that you study this manual thoroughly before operating the microscope. Retain this instruction manual in an easily accessible place near the work desk for future reference. A X 7 3 5 1
CKX41/CKX31 CONTENTS IMPORTANT — Be sure to read this section for safe use of the equipment. — 1-3 1 NOMENCLATURE 4-5 2 CONTROLS 6-7 3 SUMMARY OF OBSERVATION PROCEDURE 8 4 USING THE CONTROLS 9-15 4-1 Microscope Frame ............................................................................................................................................................................... 9 1 Turning On the Light Source 2 Adjusting the Brightness 3 Adjusting the Tension of the Coarse Adjustment Knob 4-2 Stage..................................................................................................................................................................................................................... 10 1 Placing the Specimen 2 Moving the Specimen 4-3 Observation Tube ....................................................................................................................................................................... 11-14 1 Adjusting the Interpupillary Distance 2 Adjusting the Diopter 3 Using the Eye Shades 4 Using Eyepiece Micrometer Disks 5 Selecting the Light Path (U-TR30-2 Only) 6 Adjusting the Tilt 4-4 Illumination Column........................................................................................................................................................................ 15 1 Using the Filters 2 Using the Aperture Iris Diaphragm 3 Removing the Condenser Lens 4-5 Objective Correction Collars ............................................................................................................................................. 16 1 Correction Collar 5 PHASE CONTRAST OBSERVATION 17-19 6 PHOTOMICROGRAPHY & TV OBSERVATION 20 7 TROUBLESHOOTING GUIDE 21-22 8 SPECIFICATIONS 23-24 9 ASSEMBLY — See this section for the replacement of the light bulb. — 25-29 PROPER SELECTION OF THE POWER SUPPLY CORD........................................................ 30-31
CKX41/CKX31 IMPORTANT The difference between the CKX31 and CKX41 microscopes lie in the following basis systems. CKX31 CKX41 Observation tube Binocular tube fixed Replaceable* Stage center plate – Replaceable** Reflected fluorescence system Not mountable Mountable * The U-CBI30-2/U-BI30-2/U-CTBI/CKX-TBI binocular tube or U-CTR30-2/U-TR30-2 trinocular tube can be mounted. But no intermediate attachment can be mounted. **The standard stage center plate can be replaced with the CK40-CPG glass center plate, IX-CP50 center plate (50 mm), etc. SAFETY PRECAUTIONS 1. After the equipment has been used in an observation of a specimen ? ? that is accompanied with a potential of infection, clean the parts coming in contact with the specimen to prevent infection. · Moving this product is accompanied with the risk of dropping the speci- men. Be sure to remove the specimen before moving this product. @ · In case the specimen is damaged by erroneous operation, promptly take the infection prevention measures. 2. Install the microscope on a sturdy, level table or bench (Weight: CKX31 about 8 kg, CKX41 about 8.8 kg). 3. When transporting the microscope, be sure to keep it horizontal by holding it by the root of the observation tube relay tube 1 and the illumination column 2. (Fig. 1) Fig. 1 Be sure to remove the specimen since it may fall. #Do not tilt it during carrying to prevent damaging the desktop or microscope. 4. Culture liquid or water spilt on the stage, objective or microscope frame may damage the equipment. Immediately disconnect the power cord from the wall outlet and wipe the liquid or water off if it is spilt on them. 5. The surfaces of the lamp socket 3 of the illumination column will become 5 extremely hot during operation. When installing the microscope, make sure to allow ample free space around and in particular above the lamp socket. (Fig. 1) When using the TV camera or photomicrographic system, install them so that the connection cable does not come in contact with the lamp socket. 6. To avoid potential shock hazards and burns when replacing the light 4 bulb, set the main switch 4 to “ ” (OFF) then disconnect the power cord from the wall outlet in advance. Whenever you replace the bulb during Fig. 2 use or right after use, allow the lamp socket 3 and bulb to cool before touching. (Figs. 1 & 2) Designated 6V30WHAL high-intensity halogen bulb bulbs: (PHILIPS 5761) 7. Always use the power cord provided by Olympus. If no power cord is provided, please select the proper power cord by referring to the section “PROPER SELECTION OF THE POWER SUPPLY CORD” at the end of this instruction manual. If the proper power cord is not used, product safety performance cannot be warranted. 8. Always ensure that the grounding terminal of the microscope and that of the wall outlet are properly connected. If the equipment is not grounded, Olympus can no longer warrant the electrical safety performance of the equipment. 1